N2O Wastewater Sensor Head
Expected lifetime of 6 months
Easily replaceable
Kept clean by the flow around it

N2O Wastewater Sensor Head

A Clark-type electro-chemical sensor that measures the concentration of nitrous oxide

The Sensor Head connects to the sensor body, which ensures that the cathode in the Sensor Head is polarized against its internal reference.
Within the sensor, nitrous oxide reduces on the metal surface of the cathode, which creates a current. The sensor body converts the nitrous oxide's reaction with the cathode into a signal which the Controller registers. Read more...

Read more
Concentration Ranges
N2O Wastewater Sensor Head

Low Maintenance and Easy to Replace

The N2O Wastewater Sensor Head measures the concentration of nitrous oxide from inside the sensor body. The sensor head's working range is between 0 - 1.5 mg N2O-N/L. Its detection limit is 0.005 mg N2O-N/L.

Time the delivery of sensor heads to your schedule

You will need to replace the sensor head around every 6 months, because the chemicals inside have a limited lifetime. For this reason, we recommend that you order or request to have new sensor heads delivered in accordance with your measurement schedule.

N2O WW Sensor parts_1400x800

Sensor calibration

You only need to maintain the sensor with a 10-minute calibration every second month. You can check out our instruction videos to see how you calibrate the sensor.

The sensor head will be kept clean by the friction force of the flow around it.

N2O Calibration

How to Replace the Sensor Head

When you receive a new sensor head, you will also receive replacement instructions. The tool you need to remove the old sensor head will be included in your order for the N2O Wastewater System.

We have also made this instructional video to support you when you need to replace the sensor head. We recommend that you use a hard surface and gloves when you push out the old sensor head.

thumbnail

Patent information

This product is covered by the following patents:

  • Patent issued in China: CN 104937405 A
  • Patent issued in USA: 9921178
  • Patent issued in Japan: JP6275744B2
  • Patent issued in European EP Region: EP2939012B1
  • Patent pending in several other jurisdictions.
Ordering Information
Ordering Information
Product Working Range N2O-N Detection Limit N2O-N Temperature Range Recommended for
Product
E-N2O Head SR*
Working Range N2O-N
0-1.5 mg/L
Detection Limit N2O-N
0.005 mg/L
Temperature Range
0-27°C
Recommended for
Most activated sludge systems
Product
E-N2O Head MR
Working Range N2O-N
0-9 mg/L
Detection Limit N2O-N
0.03 mg/L
Temperature Range
0-27°C
Product
E-N2O Head HR
Working Range N2O-N
0-110 mg/L
Detection Limit N2O-N
0.4 mg/L
Temperature Range
0-27°C
Product
E-N2O Head HT SR
Working Range N2O-N
0-1.5 mg/L
Detection Limit N2O-N
0.005 mg/L
Temperature Range
27-40°C
Product
E-N2O Head HT MR
Working Range N2O-N
0-9 mg/L
Detection Limit N2O-N
0.03 mg/L
Temperature Range
27-40°C
Recommended for
Deammonification and other side stream treatments
Product
E-N2O Head HT HR
Working Range N2O-N
0-110 mg/L
Detection Limit N2O-N
0.4 mg/L
Temperature Range
27-40°C

*Abbreviations: SR = Standard Range | MR = Medium Range | HR = High Range | HT = High Temperature